Waters MzML Tool

This is a CLI/GUI tool for re-calibrating Waters .mzML files, as well as fixing incorrect metadata generated when exporting .mzML files from a XevoMRT instrument.

What's wrong with Waters .mzML files?

When converting Waters .raw files into .mzML files using MSConvert, the lockmass scans are often left in the final output. This causes bugs in data processing software like MZMine: Bug often seen in .mzML files generated from data acquired on Waters instruments

It should be noted that this issue can be fixed solely using MSConvert. See these instructions.

While MSConvert can be used to remove the lockmass scans, it only supports single-point calibration. This tool also allows users to calibrate using an arbitrary number of signals, and then remove the lockmass scans (so that the files play more politely with non-Waters MS processing software)

Why re-calibrate?

Comparison of data calibrated by vendor vs calibrated using this tool Generated from an LC/MS dataset acquired on the Xevo XMRT, which is specc'd for 0.5 ppm RSME

The .mzML files exported using Waters' default software leaves a lot on the table. This only seems to be an issue with the exported data - that is, the m/z values in the .mzML files are slightly different from the values given by the native software.

This tool attempts to bridge the gap.


How to: Install

If using Windows or Linux (Debian), you can just download and launch the binary from the releases page.

Otherwise, install the repo into a virtual environment and run waters_mzml_tools.py. Instructions for building your own binaries are in BUILD.md.

How to: Retain lockmass scans when exporting from Waters instruments

This section is not exhaustive. I've tested .mzML exports on two Waters instruments: Synapt G2-Si, and Xevo MRT.

  • Obtain .raw data file (i.e. 'MassLynx format')
    • If using a Xevo MRT, this can be generated using 'DataConvert'
  • Use MSConvert with the following parameters:
    msconvert /raw/data/file.raw --filter "peakPicking vendor"
  • This generates .mzML files that still have the lockmass scans included

Note: Converting the Xevo MRT's .raw files into .mzMLs inadvertently sets all scans to be incorrectly labeled as ms_level = 1. This tool can fix that.

How to: Apply double lockmass correction

'Adjust .mzML files' Tab

  1. Click 'Add Files'/'Remove Files' and select .mzML files prepared as per the previous section. It's critical that these files still have the lockmass scans in them.
  2. (Optional), select a directory to save the modified .mzML files. If no directory is selected, the output will be saved in the same directory as the input, but with a different file name.
  3. These options control whether the lockmass scans are to be retained in the final output. If you are processing Xevo MRT data, enable 'Fix MS level metadata'
  4. Specify calibration parameters, including:
    • The path to a .txt file containing calibrant masses on each line. The masses can either be given as m/z values, or as molecular formulae (i.e C28H37N5O7H+). If using formulae, remember to include the charge.
    • The m/z tolerance and minimum intensity to use when searching for calibrants
  5. Click 'Execute' to run the program.

Inspecting calibration curves

'View Calibration Plots' tab

Whenever this program performs a calibration, a _calibration_report.json file is saved in the same directory as the .mzML file. In the 'View Calibration Plots' tab, the user can open these .json files to generate plots of the calibration performance.

CLI Usage

The progam also has a command-line interface, and can be called directly from the command-line:

$ ./waters_mzml_tools --help
usage: waters_mzml_tools [-h] [--gui] [--input INPUT] [--experiment-type {dda,dia}] [--add-ms-levels] [--output OUTPUT] [--calibrants CALIBRANTS] [--formulae] [--calibrant-window CALIBRANT_WINDOW]
                         [--calibrant-min-intensity CALIBRANT_MIN_INTENSITY] [--remove-calibration-scans] [--show-calibration-plot]

Tool for manually calibrating Waters .mzML files. Can also be used to fix mis-formed .mzML files generated by Xevo MRT instruments

options:
  -h, --help            show this help message and exit
  --gui                 Launch GUI mode. All subsequent arguments will be ignored
  --input INPUT, -i INPUT
                        Input .mzML file, or alternatively, a report .JSON file. If a .json file is passed, will display calibration plots If this argument is not given, the program launches in GUI mode.
  --experiment-type {dda,dia}
                        Type of experiment ("dda" or "dia"). Note: dda not yet implemented
  --add-ms-levels       Add MS level tags based on the spectra native IDs. This is needed because converting XMRT data into .RAW then .mzML strips out MS level information
  --remove-calibration-scans
                        Removes the "lockmass" scans from final output

Output Directory:
  --output OUTPUT       (Optional) Select directory to save output. If not provided, will save output files in same directory with a different name

Calibration:
  Calibrate .mzML file using lockmass scans

  --calibrants CALIBRANTS, -c CALIBRANTS
                        If provided, applies calibration. Specifies location of .txt file containing calibrant m/z values or formulae
  --formulae, -f        Calibrant list contains formulae rather than m/z values. If this flag is provided, the program will calculate the expected m/z values. Remember to include charge.
  --calibrant-window CALIBRANT_WINDOW, -cw CALIBRANT_WINDOW
                        The window (in Da) for which to search for a calibrant signal. Default is 0.1
  --calibrant-min-intensity CALIBRANT_MIN_INTENSITY, -ci CALIBRANT_MIN_INTENSITY
                        The minimum signal intensity at which a calibrant is considered. Default is 1e3
  --show-calibration-plot
                        Show a calibration plot after correcting the .mzML file

About