This is a CLI/GUI tool for re-calibrating Waters .mzML files, as well as
fixing incorrect metadata generated when exporting .mzML files from a
XevoMRT instrument.
When converting Waters .raw files into .mzML files using MSConvert, the
lockmass scans are often left in the final output. This causes bugs in data
processing software like MZMine:
It should be noted that this issue can be fixed solely using MSConvert. See these instructions.
While MSConvert can be used to remove the lockmass scans, it only supports single-point calibration. This tool also allows users to calibrate using an arbitrary number of signals, and then remove the lockmass scans (so that the files play more politely with non-Waters MS processing software)
Generated from an LC/MS dataset acquired on the Xevo XMRT, which is specc'd
for 0.5 ppm RSME
The .mzML files exported using Waters' default software leaves a lot on the
table. This only seems to be an issue with the exported data - that is, the
m/z values in the .mzML files are slightly different from the values given
by the native software.
This tool attempts to bridge the gap.
If using Windows or Linux (Debian), you can just download and launch the binary from the releases page.
Otherwise, install the repo into a virtual environment and run
waters_mzml_tools.py. Instructions for building your own binaries are in
BUILD.md.
This section is not exhaustive. I've tested .mzML exports on two Waters
instruments: Synapt G2-Si, and Xevo MRT.
- Obtain
.rawdata file (i.e. 'MassLynx format')- If using a Xevo MRT, this can be generated using 'DataConvert'
- Use
MSConvertwith the following parameters:msconvert /raw/data/file.raw --filter "peakPicking vendor" - This generates
.mzMLfiles that still have the lockmass scans included
Note: Converting the Xevo MRT's .raw files into .mzMLs inadvertently sets
all scans to be incorrectly labeled as ms_level = 1. This tool can fix that.
- Click 'Add Files'/'Remove Files' and select
.mzMLfiles prepared as per the previous section. It's critical that these files still have the lockmass scans in them. - (Optional), select a directory to save the modified
.mzMLfiles. If no directory is selected, the output will be saved in the same directory as the input, but with a different file name. - These options control whether the lockmass scans are to be retained in the final output. If you are processing Xevo MRT data, enable 'Fix MS level metadata'
- Specify calibration parameters, including:
- The path to a
.txtfile containing calibrant masses on each line. The masses can either be given as m/z values, or as molecular formulae (i.eC28H37N5O7H+). If using formulae, remember to include the charge. - The m/z tolerance and minimum intensity to use when searching for calibrants
- The path to a
- Click 'Execute' to run the program.
Whenever this program performs a calibration, a _calibration_report.json file is saved in the same
directory as the .mzML file. In the 'View Calibration Plots' tab, the user can open these .json files
to generate plots of the calibration performance.
The progam also has a command-line interface, and can be called directly from the command-line:
$ ./waters_mzml_tools --helpusage: waters_mzml_tools [-h] [--gui] [--input INPUT] [--experiment-type {dda,dia}] [--add-ms-levels] [--output OUTPUT] [--calibrants CALIBRANTS] [--formulae] [--calibrant-window CALIBRANT_WINDOW]
[--calibrant-min-intensity CALIBRANT_MIN_INTENSITY] [--remove-calibration-scans] [--show-calibration-plot]
Tool for manually calibrating Waters .mzML files. Can also be used to fix mis-formed .mzML files generated by Xevo MRT instruments
options:
-h, --help show this help message and exit
--gui Launch GUI mode. All subsequent arguments will be ignored
--input INPUT, -i INPUT
Input .mzML file, or alternatively, a report .JSON file. If a .json file is passed, will display calibration plots If this argument is not given, the program launches in GUI mode.
--experiment-type {dda,dia}
Type of experiment ("dda" or "dia"). Note: dda not yet implemented
--add-ms-levels Add MS level tags based on the spectra native IDs. This is needed because converting XMRT data into .RAW then .mzML strips out MS level information
--remove-calibration-scans
Removes the "lockmass" scans from final output
Output Directory:
--output OUTPUT (Optional) Select directory to save output. If not provided, will save output files in same directory with a different name
Calibration:
Calibrate .mzML file using lockmass scans
--calibrants CALIBRANTS, -c CALIBRANTS
If provided, applies calibration. Specifies location of .txt file containing calibrant m/z values or formulae
--formulae, -f Calibrant list contains formulae rather than m/z values. If this flag is provided, the program will calculate the expected m/z values. Remember to include charge.
--calibrant-window CALIBRANT_WINDOW, -cw CALIBRANT_WINDOW
The window (in Da) for which to search for a calibrant signal. Default is 0.1
--calibrant-min-intensity CALIBRANT_MIN_INTENSITY, -ci CALIBRANT_MIN_INTENSITY
The minimum signal intensity at which a calibrant is considered. Default is 1e3
--show-calibration-plot